MODEL 3548COD High Temperature
Furnace COD Gages (1200 °C or 1600 °C)

High temperature COD gages for use in split type materials testing furnaces with a slot for the ceramic rods. Water-cooled and furnace bracket mounted, these extensometers are for use to 1200 °C (2200 °F). The high temperature option allows use to 1600 °C (2900 °F). They are specifically designed for fracture mechanics testing.

These COD gages mount on a water-cooled bracket, which is mounted on the furnace side cut-out or with optional load frame support brackets. The standard temperature version (to 1200 °C) is supplied with high purity alumina rods. The high temperature option is furnished with alpha grade silicon carbide rods. Rods are made to order to the length required for your furnace. Mounting brackets may be integrated with the furnace cut-out. Epsilon can also provide optional load frame mounting brackets to fit your test frame. Contact edges on the test sample should be somewhat rounded (not sharp knife edges) for best performance with this model.

The Model 3548COD extensometers are strain gaged devices, making them compatible with any electronics designed for strain gaged transducers. Most often they are connected to a test machine controller. The signal conditioning electronics for the extensometer is typically included with the test machine controller or may often be added. In this case the extensometer is shipped with the proper connector and wiring to plug directly into the electronics. For systems lacking the required electronics, Epsilon can provide a variety of solutions, allowing the extensometer output to be connected to data acquisition boards, chart recorders or other equipment.

Features

  • Full bridge, 350 ohm strain gaged design for compatibility with nearly any test system.
  • All standard units meet existing ASTM requirements for accuracy.
  • All units come with either high purity alumina ceramic rods (1200 °C) or alpha grade silicon carbide rods (1600 °C).
  • Rugged, dual flexure design for strength and improved performance.
  • Includes high quality foam lined case and a spare set of ceramic rods.

 

Model 3548COD Available Versions: ANY combination of gauge length, measuring range and temperature range listed below is available, except as noted. Other configurations may be available with special order; please contact Epsilon to discuss your requirements.

3548COD extensometer chart

Cemark logo  CERTIFIED

Specifications
Excitation: 5 to 10 VDC recommended, 12 VDC or VAC max.
Output: 2 to 4 mV/V, nominal, depending on model
Linearity: ≤0.15% of full scale measuring range, depending on model
Temperature Range: Standard (-ST) is to 1200 °C (2200 °F), optional (-HT) 1600 °C (2900 °F)
Cable: Integral, ultra-flexible cable, 8 ft (2.5 m) standard
Operating Force: <30 g typical

 

Options
  • Load frame mounting brackets
  • High temperature option (-HT suffix) for use to 1600°C
  • Connectors to interface to nearly any brand test equipment

 

> Shunt Calibration

 

 

contact us


 

 

click below for larger view
Model 3548COD mounted to a compact tension specimen
Model 3548COD mounted to a compact tension specimen

Model 3548 with an open furnace showing specimen engagement
Model 3548COD with no heat shields mounted to a Model 3590VHR calibrator